Back to Top

Hi, Guest!

  LOKASI :  Kota Bekasi

Bergabung Selama :

BAGIKAN :   

Bagikan :
  • Analitik
  • Single Beam UV/Vis Spectrometer DU-8200

Single Beam UV/Vis Spectrometer DU-8200

Update Terakhir
:
29 / 10 / 2024
Min. Pembelian
:
1 Unit
Dilihat Sebanyak
:
69 kali

Harga

CALL
Bagikan
:
PT. Fedito Bahtera Abadi adalah perusahaan supplier alat laboratorium. Produk yang bisa kami suplai konsumabel laboratorium, peralatan laboratorium dan alat ukur.
Dengan didukung tenaga aftersales service yang berpengalaman, kiranya kami bisa menjadi partner untuk pengadaan alat laboratorium di Indonesia.

Perhatian !

Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.

Detail Single Beam UV/Vis Spectrometer DU-8200

Spectrometer

Features:
1. Small fine design, space-saving
2. The excellent optical and electric system
3. Low noise and Low stray light
4. Large LCD display (128*64 Dots), can display curve
5. High-quality grating, detector and lamps
6. Data and Curve can be stored in real-time
7. Auto setting WL, auto Blank
8. Easy to change Pri-aligned lamps
9. Reinforced baseboard and bracket assures
Function:
1. Photometric: T%, Abs
2. Quantitative: Standard Curve
3. System Utility
4. WL Scan (Spectrum Scan) (PC)
5. Time Scan (Kinetics) (PC)
6. DNA/Protein Test (PC)
7. Multi-WL Test (PC)Specifications of the Visible Spectrophotometer
Model
DU-8200
Wavelength Range
190-1100nm
Spectral Bandwidth
2.0nm or 4.0nm optional
Optical System
Single Beam, Blazed Holographic Grating (1200 lines/mm)
Wavelength Accuracy
±0.8nm
Wavelength Repeatability
≤0.2nm
Wavelength Setting
Auto, Resolution 0.1nm
Photometric Range
0~200%T, -0.3­~3A, 0~9999C
Photometric Accuracy
±0.002A (0~0.5A), ±0.004A (0.5~1A), ±0.3%T (0~100%T)
Photometric Repeatability
≤0.001A (0~0.5A), ≤0.002A (0.5~1A), ≤0.2%T (0~100%T)
Stray Light
≤0.1%T(220/360nm)
Scan Speed
High, Medium, Low. Max.1000nm/minute
Baseline Flatness
±0.003A
Stability
±0.002A/h (500nm,0A)
Noise
≤0.2%T/3min (250/500nm,0%T); ≤0.5%T/3min (250/500nm,100%T)
Sample Compartment
Accommodate 5-100mm Pathlength Cuvette
Detector
Silicon Photodiode
Lamps
Tungsten Lamp & Deuterium Lamp (Pre-aligned)
Display
Graphic LCD (128×64 Dots)
Keypad
11-key Membrance Keypad
Output Port
USB Port & RS232 Port
Printer
Mini Serial Printer; PC Printer
PC Software
Optional PC Scanning Software
Power Requirements
AC 90-250V,50/60Hz
Dimension
431x318x195mm
Net Weight / Gross Weight
9kgs / 13kgs